Testing new probe with new pulser
Experiment 20171112a
Context
Just got a new pulser and a test probe (ATL again): let's test those!
- Using a streamlined conversion, unpacking and display script
- Using a image tagging script
- Jupyter notebook of the test
- On the Pi, the kernel module is as follows
Unpacking acquisitions:
TLDR: the new probe seems to be rotating twice as fast, and gives no echoes. I bet this is due to a lack of fluid inside the probe (as the piezo still seem to exist).
New probe
Old probe
Into spectra
TLDR: the new probe seems to have a frequency lower of that of the probe I had. Could that be due to the lack of fluid ?
New probe
Old probe
Unpacking and SC'ing the images
Also worked on the lib to get streamlined images scan converted straight from the binary files. It helps!